X-ray HR CCD cameras
Price per week under a 3 years contract
Free Shipping in Canada and U.S.
Buy now, get in 10 days
- Small pixel size less < 26 microns pixel size at the detector input
- Cooled CCD sensor with 55 degrees C delta T
- Software switchable 10 and 20 MHz scanning frequencies
- Small area sensor with taper input varying from 5.6 mm up to 40 mm diagonal
PSL has supplied X-ray High Resolution CCD cameras for the last 7 years to end users and OEMs. A selection of high responsitivity CCDs, combined with low noise characteristics, enables optimum photonic collection with best possible signal to noise ratio. Special read whilst expose mode allows 100% shutterless duty cycle and high sensitivity operation in low light level conditions.
Applications:
- • Microdiffraction
- • X-ray imaging
- • X-ray micro CT
- • Laue imaging
- • Protein crystallography at up to 50 keV
- • Gisaxs
- • Powder Diffraction
- • Non Destructive Testing
- • Phase Contrast Imaging
- • Small animal imaging
X-ray megapixel HR CCD camera
- 1392 (h) x 1040 (v) CCD array
- Input pixel size : available from 3.2 x 3.2; 6.45 x 6.45; 10.37 x 10.37; 26 x 26 and up to 31.18 x 31.18 microns
- Input size : available from 18 x 12 mm; 36 x 24 mm; 75 x 50 mm; 14.4 x 14.4 mm and 23.05 x 23.05 mm respectively
- 11 fps at full resolution @ 20 MHz and 5 fps @ 10 MHz
- Readout noise : 4 - 5 electrons @ 10 MHz with interpolation noise reduction
- Full well capacity : 13,000 electrons in binning 1x1; 22,000 electrons in binning 2x2
- Dark current : < 0,01 electrons / pixel / second
- 12-bit digitisation
- 16-bit extended dynamic range
- GdOS:Tb scintillator for operation from 5-55 keV with minimum feature recognition of 10 lp / mm: typically 12 microns for the smallest input size up to 50 microns for the largest input size.
- CsI:Tl structured scintillator for operation from
- 30-100 keV
- Camera link / GigE interface
- Synchronisation / control : via TTL pulse
X-ray 4 megapixel HR CCD camera
- 2048 (h) x 2048 (v) CCD array
- Input pixel size : available from 3.8 x 3.8; 7.4 x 7.4; 17.3 x 17.3; 25.9 x 25.9 microns
- Input size: available from 7.8 x 7.8 mm; 15.2 x 15.2 mm; 35.4 x 35.4 mm; 53.04 x 53.04 mm respectively
- 2 fps at full resolution @ 12 MHz
- 3.5 fps in binning 2x2 @ 1024x1024 resolution
- Readout noise: 7 - 9 electrons @ 12 MHz with interpolation noise reduction
- Full well capacity : 40,000 electrons in binning 1x1; 80,000 electrons in binning 2x2
- Dark current : 0,05 electrons / pixel / second
- 16-bit extended dynamic range
- USB 2.0 interface
- Synchronisation / control : via TTL pulse
Additional Features:
- Low readout noise < 7 electrons with noise interpolation reduction
- Very low dark current with less than 0.01
- Gating time from milliseconds to > 30 minutes
- Simultaneous integration / readout enabling 100% duty cycle acquisition
- GdOS polycrystalline or structured CsI scintillators
- On chip binning
- Camera link and GigE digital interface
- Peltier / fan cut off option
- C-mount
Durham university - HR X-Ray document
X-ray HR camera application note
Vacuum Flange